Audible signal function, setting saved
Note: If the audible signal function is activated, the timer is indicated by a signal
sequence (3 tones in 3 seconds).

Setting the lock function

Note: All devices with the abbreviation „NOL" in the model name (power rating
plate) do not have a lock function.
The locking function for the shaking table can be activated or deactivated directly
after the setting of the acoustic signal.
The following steps are required to activate or deactivate the locking function:
a.) Turn off the device with the power switch.
b.) Press and hold the right rotary knob and switch the device on with the
power switch.
After approx. 2 seconds, the display shows "bEEP on" or "bEEP oFF"
(depending on the previous setting) and the rotary knob can be released.
Confirm the settings or wait until the display "Lock on" or "Lock oFF"
(depending on the previous setting).
c.) By pressing the right rotary knob, as long as the LED of the set indicator flashes,
the locking function can be activated (Lock on) or deactivated (Lock oFF).
Lock function activated
24
Locking function deactivated
Note: If no changes are made in 3 seconds, the setting is saved and the LED of
the set indicator stops flashing.
Locking function, setting saved
After that, the overall operating time (hours) appears on the screen.
Set
rpm
Operating hours, for example: 12345 hours
The activated locking function is indicated by the illuminated LED of the status
indicator Lock function is displayed.
ATTENTION! After switching on the device with the power switch
!
or after a power failure (equivalent to disconnecting from the power),
the device searches for the locking position by a defined movement
of the shaking table and fixes it.