Operating modes

The device can be operated in the modes described below:
Operation mode A (factory setting):
After switching on the device with the power switch, the shaking and timer
functions are switched off and the stored set value are displayed. The set value can
be changed.
After a power failure (equivalent to disconnecting from the power), the device
does not restart by itself in mode A. Previously functions must be restarted.
The upper speed limit can only be set or changed in operating mode A.
Operating mode B:
After switching on the device with the power switch, the shaking and timer
functions are switched off and the stored set values are displayed. The set
values can be changed.
When switching to operating mode B, the setting of the upper speed limit is
taken over from operating mode A and cannot be changed.
After a power failure (equivalent to disconnecting from the mains), the device
does not restart itself in mode B. Previously functions must be restarted.
Operating mode C:
When switching to operating mode C, the set values for: upper speed limit,
target speed and timer are taken over from operating mode B and cannot be
changed.
After switching on the device with the power switch or after a power
!
failure (equivalent to disconnecting from the mains), the device starts
up again in mode C if the shaking function was activated previously.
The stored set values are displayed accordingly.
Switching the operating mode:
The following steps are required to switch the operating modes:
a.) Turn off the device with the power switch.
b.) Press and hold both rotary knobs and turn on the device with the power switch.
After about 1 second, the rotary knobs can be released.
Note: It is not possible to switch between operating modes A, B and C in
random order. The operating mode can be switched by repeating these two
steps a.) and b.) according to the following steps.
A - step a.) + b.) => B - step a.) + b.) => C - step a.) + b.) => A - step a.) + b.) =>

Setting the audible signal function

To activate or deactivate the audible signal function, the following steps are necessary:
a.) Turn off the device with the power switch.
b.) Press and hold the right knob and turn on the device with the power switch.
After approx. 2 seconds, the display shows "bEEP on" or "bEEP off"
(depending on the previous setting) and the rotary knob can be released.
c.) By pressing the right rotary knob, the signal function can be activated (bEEP on)
or deactivated (bEEP off) as long as the LED of the set indicator flashes.
When switching from "bEEP off" to "bEEP on", a sound signal sounds.
Audible signal function activated
Audible signal function deactivated
Note: If no changes are made in 3 seconds, the setting will be saved and the set
indicator LED will stop flashing.