ADJUSTMENT
5) Put the set into the adjustment mode.
•Turn OFF the power.
•Press both the "+" and "Density" keys.
•Release the "+" and "Density" keys.
•Press the "Color" key.
•Confirm that the numeric LED display is 00.
•Press the "Repeat" key.
•The florescent lamp lights after four seconds and the set enters the CCD adjustment mode.
•Press the "Repeat" key.
•Confirm that the fluorescent lamp lighted.
•Confirm that the numeric LED display is 54.
6) Adjustment of oscilloscope (See Figure 3.)
Adjust the trigger level and Ch1 vertical position so that a CCD waveform appears on the oscilloscope display.
Adjust the horizontal position, luminance, and focus.
Fig.3