Filling - Julabo DYNEO DD-BC4 Operating Instructions Manual

Immersion/open bath/refrigerated circulators
Table of Contents

Advertisement

1.888.610.7664

Filling

Danger from damaged safety devices
Possible serious consequences for personnel and working areas.
Check the safety devices at least twice per year.
Excess temperature protector, IEC 61010-2-010
Turn the adjustable excess temperature protector to the cut-out point
(actual temperature) using a screwdriver. The actors will be shut down
+
+
on all poles, the circulator will show error message E 14, the "Alarm"
control display will be lit and a continuous signal tone will sound.
E 14
Low level safety device, IEC 61010-2-010
The float switch in this device must be moved manually in the bath to
test the function, for example using a screwdriver.
Push down the float until its reaches the mechanical stop.
The actors will be shut down on all poles, the circulator will show error
message E 01, the "Alarm" control display will be lit and a continuous
signal tone will sound.
Pos: 3 1 /Technis che D okumentation/Flash CORI O/ CORIO_ CD_GESAMT /Füllen - Grundlegende Ge fahren @ 0\ mod_1409 2184 4925 1_18. docx @ 6708 @ 1 @ 1
10 Filling
Basic dangers.
The volume of oil used as bath fluid changes with the temperature.
Starting from the volume when the bath is filled (room temperature) it
may increase or decrease during operation.
The bath temperature rises - hot bath fluid can overflow.
The bath temperature falls - the low level alarm will stop the tempering
process.
Monitor the level until it reaches working temperature.
Filling process
Ensure that the drain valve is closed. Turn the knurled screw.
Carefully insert bath fluid - never allow bath fluid to get inside the
circulator.
Do not exceed the maximum bath capacity (see Technical Data).
1 - The bath temperature rises - hot bath fluid can overflow.
2 - The bath temperature falls - the low level alarm will stop the
tempering process.
Monitor the level until it reaches working temperature.
Pos: 3 2 /Technis che D okumentation/Flash CORI O/ CORIO_ CD_GESAMT /Pumpe neinstell ung @ 0 \mod_1 40921 8452 964_1 8.docx @ 68 73 @ 2 @ 1
www.calcert.com
15
10
20
5
25
0
30
29