Commissioning; Excess Temperature And Low Level Safety Devices - Julabo DYNEO DD-BC4 Operating Instructions Manual

Immersion/open bath/refrigerated circulators
Table of Contents

Advertisement

1.888.610.7664
9

Commissioning

Danger from mains voltage.
Risk of injury from electric power.
Commissioning the circulator with a refrigeration machine
Connect the circulator and refrigeration machine using the mains lead.
Connect them to the voltage supply using the fitted plug on the
refrigeration machine and the mains lead. Connect the CAN jacks on
both devices with the CAN connection cable to transfer data.
Pos: 2 9 /Technis che D okumentation/Flash CORI O/ Warnu ng heiße Ober fläche @ 0\ mod_13 8968 60883 20_18 .docx @ 366 8 @ @ 1
Cold or hot unit surfaces
Risk of frostbite, risk of burns
What should be observed when operating the JULABO
temperature control unit?
Pos: 3 0.1 /Te chni sche Dokumentation/Flash CORIO/S chutzeinrichtung, Überte mperatur, U nterniveau - Erklärung @ 0\ mod_1 38968 8531 828_1 8.docx @ 372 8 @ 2 @ 1
9.1

Excess temperature and low level safety devices

The safety devices are not affected by the control circuit. When they trip
all actors are permanently shut down.
The alarm is displayed optically and acoustics with a continuous signal
tone and the reason for the alarm is shown on the display as a number.
Pos: 3 0.2 /Te chni sche Dokumentation/Flash CORIO/ Überte mperatur -, U nterniveau-Sch utzeinri chtung -War nung @ 0\ mod_ 1389 68859 9281_ 18.docx @ 3 734 @ @ 1
28
www.calcert.com
15
10
20
5
25
0
30
Compare the mains voltage and frequency with the details on the
model plate.
Connect the device only to a safe power supply via FI-circuit
breaker (IA = 30 mA).
The device may only be connected to power outlets with a ground
contact (PE – protective earth).
The mains plug serves as a safe disconnecting device from the
power supply network and must be freely accessible at all times.
Do not start the device if it has a damaged mains cable.
Check the mains cable regularly for signs of damage.
We disclaim all liability for damage caused by incorrect line
voltages!
Unit parts may develop high surface temperatures. A hot surface
means it has a temperature of 60 °C / 140 °F or more.
Let the device cool down to an uncritical safe temperature.
Use safety gloves.