6. Practical Use
Step 5:
Start the program.
Activate the program by pressing the START key. The LED lights up in
green. The progress of the program is shown in the display.
The furnace head opens automatically at the end of the program.
The results panel displays the press time, press distance and the
investment ring temperature, if the IRT investment ring analysis is
activated, at the time of the loading.
After confirmation of the results panel and as soon as the Optical
Status Display lights up in green, the furnace is ready for the next
program start.
6.4 Press procedure by means of FPF – fully automatic press function
Step 1:
Switch the furnace on.
Switch the furnace on using the mains switch at the back of the
furnace. Subsequently, the furnace conducts an automatic self-test.
Wait until the furnace has reached the pre-set stand-by temperature.
Step 2:
Select the operating mode.
After the self-test has been completed, the selection of the operating
modes is displayed. Select the operating mode "Pressing".
78