Warranty /
Procedure in case of repair
The warranty period for Bluephase N is 3 years from the
date of purchase (battery 1 year).
Malfunctions resulting from faulty material or
manufacturing errors are repaired free of charge during
the warranty period. The warranty does not provide the
right to recover any material or non-material damage
other than the ones mentioned. The apparatus must only
be used for the intended purposes. Any other uses are
contraindicated. The manufacturer does not accept any
liability resulting from misuse and warranty claims
cannot be accepted in such cases. This is particularly
valid for:
– damage resulting from improper handling, especially
incorrectly stored batteries (see Technical data:
Transportation and storage conditions).
– damage to components resulting from wear under
standard operating conditions (e.g. battery).
– damage resulting from external influences, e.g. blows,
drop to the floor.
– damage resulting from incorrect set-up or installation.
– damage resulting from connecting the unit to a power
supply, the voltage and frequency of which do not
comply with the ones stated on the rating plate.
– damage resulting from improper repairs or modifica-
tions that have not been carried out by certified
Service Centers.
In case of a claim under warranty, the complete
apparatus (handpiece, charging base, power cord, and
power pack) must be returned, carriage paid, to the
dealer or directly to Ivoclar Vivadent, together with the
purchase document. Use the original packaging with the
corresponding cardboard inserts for transportation.
Repair work may only be carried out by a certified
Ivoclar Vivadent Service Center. In case of a defect that
cannot be rectified, please contact your dealer or your
local Service Center (see addresses on the reverse side).
A clear description of the defect or the conditions under
which the defect occurred will facilitate locating the
problem. Please enclose this description when returning
the apparatus.

Product specifications

Delivery form
1 Charging base, 1 Power cord, 1 Power pack, 1 Handpiece, 1 Battery
(Li-Po), 1 10 mm light probe, 3 Anti-glare cones, 1 Set of Instructions
for Use
Accessories
The following accessories are available for Bluephase N:
REF
Description
656196
Light probe 10 mm, black (N)
608538
Pin-point light probe 6>2 mm, black
551756
Anti-glare cones
592496
Anti-glare shield
659956
Bluephase N battery
659681
Bluephase N handpiece
607922
Bluephase Meter
608554
Bluephase sleeves
Technical data
Operating voltage charging base
Operating voltage handpiece
Power pack
Output
TYPE 15.2630
Operating conditions
Temperature
Relative humidity
Ambient pressure
Dimensions of the charging base
Weight of the charging base
Charging time
Power supply of the handpiece
Max. battery time
Light source
Wavelength range
Light intensity
Operating
Dimensions of the handpiece:
Weight of the handpiece
Transportation and storage conditions:
Temperature
Relative humidity
Ambient pressure
Bluephase N has to be stored in closed, roofed rooms.
Protect the device from severe jarring.
Battery
– Do not store at temperatures above 40 °C / 104 °F
(or 60 °C / 140 °F for a short period).
Recommended storage temperature 15 – 30 °C (59 – 86 °F).
– Always store the battery charged
– and not for longer than 6 months.
5 VDC
3.7 VDC with battery
5 VDC with power pack
100–240 VAC / 50-60 Hz /
max. 0.4 A
Manufacturer: Friwo
5 VDC
+10 °C to +35 °C (+50 to +95 °F)
30% to 75%
700 hPa to 1060 hPa
L=205mm W=150mm H=85mm
250 g
approx. 2 h (with the battery empty)
Li-Po battery
approx. 60 min. (with a new, fully
charged battery)
Polywave
LED
®
385 – 515 nm
max. 1,200 mW/cm
± 10 %
2
5 min. on / 6 min. off
(intermittently)
L=260mm W=42mm H=120 mm
(incl. battery) 225 g
-20 °C to +60 °C (-4 to 140 °F)
10% to 75%
500 hPa to 1060 hPa