Cleaning; Service Note; Stand-By; Power-Saving Mode - Ivoclar Vivadent 747832BU Operating Instructions Manual

Table of Contents

Advertisement

7. Maintenance, Cleaning and Diagnosis

7.2 Cleaning

The furnace may only be cleaned when it is cool, since there is a burn hazard. Do not use any cleaning solutions.
Disconnect the power before cleaning.
The following parts have to be cleaned from time to time:
What
Furnace housing and furnace head
Membrane-sealed keypad and display
Cooling tray
Insulation
Sealing rims of the furnace head and the sealing
surface
Firing plate
Infrared camera
*Never clean with compressed air!

7.3 Service note

When the service note appears for the first time, the furnace has already more than 1,500 firing hours. If the service note is
acknowledged, it will reappear every 1,000 firing hours.
Ivoclar Vivadent recommends conducting a heater test after a certain number of firing hours and having the heating muffle
examined by the Ivoclar Vivadent After Sales Service if necessary.

7.4 Stand-by

We recommend keeping the furnace head closed, particularly if the temperature drops below 150 °C. If the furnace head
is open, there is a risk that the insulation absorbs humidity and condensation water develops during firing. This negatively
affects the vacuum performance and, consequently, the firing results.

7.5 Power-saving mode

If the furnace is not used for longer periods of time, Ivoclar Vivadent recommends to activate the power-saving mode. With
the power-saving mode activated, individual components are switched off and the furnace temperature is reduced to
100 °C.
7.5.1 Automatic power-saving mode
The automatic power-saving mode can be activated in the Settings menu.
If the power-saving mode is activated and the furnace head closed, this function is automatically started after 30 minutes,
if the furnace is in idle mode and no key is pressed during that time. The display shows the power-saving icon. Pressing any
key ends the automatically activated power-saving function.
7.5.2 POWER SAVING key
The power-saving mode can be manually activated by pressing the POWER SAVING key. This is only possible with the
furnace head closed and the furnace on stand-by. The display shows the power-saving icon. Pressing any key ends the
automatically activated power-saving function.
The power-saving mode is not available in the press operating mode.
84
When
If required
Soft, dry cloth
If required
With a Programat cleaning cloth
Daily
Cleaning brush*
Daily
Cleaning brush*
Daily
Cleaning brush and a soft cloth
If required
Cleaning brush or sandblasting unit
(with glass beads)
If required
With a Programat cleaning cloth
Cleaning material

Advertisement

Table of Contents
loading

This manual is also suitable for:

Programat ep 5010 g2645984eu

Table of Contents