Setting Up the Avant 4000 Digital Pulse Oximetry System
Use the following procedure to set up the Avant 4000 Digital Pulse Oximetry
System. (Refer to the Avant 4100 Patient Module instruction insert for additional
information about using that product.)
1.
Ensure that batteries are installed in the display unit and wrist-worn patient
module.
WARNING: The battery pack must be installed at all times while the
device is operating—even when operating on AC power. If it
is necessary to operate the device without batteries, audible
alarms and memory functions may not be available. DO NOT
use the device without batteries when patient safety relies
upon audible alarms.
2.
Plug in and connect the AC adapter for the display unit.
3.
Attach a sensor to the patient module. (The patient module is automatically
activated when a sensor is connected.)