Nonin Avant 4100 Operator's Manual page 43

Digital pulse oximetry system with bluetooth wireless technology
Table of Contents

Advertisement

T r o u b l e s h o o t i n g
Problem
A dash (-) appears in
the %SpO
display.
2
An error code appears
in the display area.
40
Possible Cause
A poor signal from the
digit is being detected.
The digit was removed
from the sensor.
The system is not
functioning.
The system encountered
an error.
Possible Solution
Reposition the digit or insert a
different digit and keep the
sensor motionless for at least
10 seconds.
Position the sensor at a
different site.
Reinsert the digit and keep the
sensor motionless for at least
10 seconds.
Turn off the unit, check all
connections, and retry.
Verify that the 4100 is paired
with the display.
Verify that the Connection
Status LED is green on both
the display unit and the patient
module.
Contact your distributor or
NONIN Customer Support.
1. Turn the unit off and then
back on again to remove
the error code.
2. If the error persists,
disconnect all power (AC
and battery), and then
reconnect the power and
turn the unit back on.
3. If the error still persists,
note the error code and
contact your distributor or
NONIN Customer
Support.