To Verify Ei System Performance - Agilent Technologies 5977C Series Operating Manual

Table of Contents

Advertisement

4
Operating in EI Mode

To Verify EI System Performance

To Verify EI System Performance

Materials needed for an EI HES
Materials needed for an EI XTR, SS, or Inert source
Verify the tune performance
1 Verify that the system has been pumping down for at least 60 minutes.
2 Set the GC oven temperature to 150 °C and the column flow to 1.0 mL/min.
3 In the Instrument Control view, select Checkout Tune from the Checkout
4 When the autotune has completed, save the method, and then select
Verify the sensitivity performance
1 Set up to inject 1 µL of the checkout sample shown in the Materials needed
2 In the Instrument Control view, select Sensitivity Check from the Checkout
3 If necessary, resolve your hardware with this method and place the sample in
4 Click OK to run the method.
When the method is completed, an evaluation report will be printed.
Verify that rms signal-to-noise ratio meets the published specification. See the
Agilent Web site at www.agilent.com/chem for specifications.
104
10 fg/µL OFN Checkout standard 3 × 1 mL (5190-0585)
1 pg/µL OFN EI Checkout standard 3 × 1 mL (5188-5348)
menu. The software will perform an autotune and print the report.
Evaluate Tune from the Checkout menu.
The software will evaluate the last autotune, and print a System Verification
– Tune report.
for your source type above, either with the ALS or manually.
menu. The system displays an Alert dialog reminder about resolving the
OFN_SN method and placing the OFN sample in vial 1 when an ALS is
configured.
the vial 1 position.
5977C GC/MSD Operating Manual