Troubleshooting
Code
21624 or
xxxxx Supply: Film Calibration
23624
Failure
21631 or
xxxxx Supply: Film Calibration
23631
Failure - Dmin Outside Target
21632 or
xxxxx Supply: Film Calibration
23632
Failure - Dmax Outside Target
25922
RF Tag: Internal Diagnostic
Failure
25931
RF Tag: Internal
Communications Failure
26325
Film Jam in Area 2
26326
Film Jam in Area 2 or 3
26543
Film Jam in Area 3
26544
Film Jam in Area 3
26931
Film Transport: Internal
Communications Failure
26933
Film Transport: Internal
Communications Failure
27123
Optics: Internal Hardware
Failure
27601
Optics: Calibration Failed
3-10
Table 3-5: Condition Codes (Cont'd)
Web Portal Message
User Action
1. Try calibration again, or try a different film pack.
2. If the error repeats, shut down the Imager, then start
the Imager.
3. If the error persists, call for service.
1. The minimum density of the film is too high.
Calibration results for this film are outside the
normal range, but will still be used.
2. If you do not wish to use these calibration results,
perform a new calibration on this film or try
another film pack.
1. The maximum density of the film is lower than the
target density. Calibration results for this film are
outside the normal range, but will still be used.
2. If you do not wish to use these calibration results,
perform a new calibration on this film or try
another film pack.
1. Shut down the Imager, then start the Imager.
2. If the error persists, call for service.
1. Shut down the Imager, then start the Imager.
2. If the error persists, call for service.
See
"Film Jam - Code
26325".
See
"Film Jam - Codes 26326 or
See
"Film Jam - Codes 26326 or
See
"Film Jam - Code
26544".
1. Shut down the Imager, then start the Imager.
2. If the error persists, call for service.
1. Shut down the Imager, then start the Imager.
2. If the error persists, call for service.
1. Shut down the Imager, then start the Imager.
2. If the error persists, call for service.
1. Shut down the Imager, then start the Imager.
2. If the error persists, call for service.
2G0733
26543".
26543".