Error codes

Any malfunctions during operation will be identified by an error message on the display�
Proceed as follows in such cases:
› Turn off the device by using the power switch�
› Carry out corrective measures�
› Restart the device�
E10412324
Causes
Temperature inside device (PCB power) too high
Effect
Tempering function off
Solutions
Switch off device and allow it to cool down�
E10393524
Causes
Set temperature limit exceeded
Effect
Tempering function off
Solutions
Contact IKA service department�
E10322713 / 10332713
Causes
Safety relay does not open
Effect
Tempering function off
Solutions
Contact IKA service department�
E10144420
Causes
Heater1 error
Effect
Tempering function off
Solutions
Contact IKA service department�
E10412923
Causes
No internal sensor
Effect
Tempering function off
Solutions
Contact IKA service department�
If the action described fails to resolve the fault or another error code is displayed then take
one of the following steps:
› Contact the service department�
› Send the device for repair, including a short description of the fault�
50

Maintenance and cleaning

The device is maintenance-free� It is only subject to the natural wear and tear of components and
their statistical failure rate�
/// Cleaning
For cleaning disconnect the mains plug!
Use only cleaning agents which have been approved by IKA to clean the devices:
These are: water (containing surfactant) and isopropyl alcohol�
› Wear protective gloves during cleaning the devices�
› Electrical devices may not be placed in the cleansing agent for the purpose of cleaning�
› Do not allow moisture to get into the device when cleaning�
› Before using another than the recommended method for cleaning or decontamination, the
user must ascertain with that this method does not destroy the device�
/// Decontamination
› The device allows hot air decontamination at 120°C for 2 hours� For details, refer to
"Decontamination" in section "Menu details"�
Note: Before decontamination, the device must be completely emptied�
/// Ordering spare parts
When ordering spare parts, please indicate:
› Device type�
› Serial number, see type plate
› Item and designation of the spare part, see: www.ika.com, spare parts diagram and spare
parts list�
› Software version (Briefly visible in the display when the device is switched on)�
/// Repairs
Please send in device for repair only after it has been cleaned and is free from any
materials which may constitute a health hazard.
For repair, please request the "Decontamination Certificate" from IKA, or download printout
of it from the IKA website www.ika.com�
If you require servicing, return the device in its original packaging� Storage packaging is not
sufficient� Please also use suitable transport packaging�

Accessories

IO T 1.10
Enhanced wire grid tray
IO F 1.10
Roller frame
IO B 1.10
Roller box
IO P 1.10
Table top work plate
See more accessories on www.ika.com�