Wait for about one to two hours until temperature is stable (e�g� reach about 24�8 °C), the "set"
symbol appears and blinks� Meanwhile, a "beep" sound is heard�
Press the control knob (H), the "key" symbol appear:
Now, you can input the calibration value from the reference measuring device (e�g� 24�9 °C) by
rotating the control knob (H)�
Press the control knob (H) to finish the first point calibration�
Then, the second point calibration screen appears automatically�
The calibration can be performed in the same steps�
46

Interfaces and outputs

The device can be connected to a PC and operated with the laboratory software labworldsoft
through USB interface�The device software can also be updated with a PC via the USB port�
Note: Please comply with the system requirements together with the operating instructions
and help section included with the software�
/// USB interface
The Universal Serial Bus (USB) is a serial bus for connecting the device to the PC� Equipped with
USB devices can be connected to a PC during operation (hot plugging)� Connected devices and
their properties are automatically recognized�
/// USB device drivers
First, download the latest driver for IKA devices with USB interface from:
www.ika.com/ika/lws/download/usb-driver.zip.
Install the driver by running the setup file� Connect the IKA device through the USB data cable
to the PC� The data communication is via a virtual COM port�
Note: the USB driver is unnecessary for Windows 10 system� Don't install it for Windows 10 system!
®