Error Codes - IKA HB digital Manual

Hide thumbs Also See for HB digital:
Table of Contents

Advertisement

Error codes

The fault is shown by an error code on the display as following if the error occurs�
Proceed as follows in such cases:
› Turn off the device by using the main switch�
› Carry out corrective measures�
› Restart the device�
Error code | Causes | Effect | Solutions
E 2
Causes
In remote operation (PC), no communication between RV 10 and HB digital
IR connection to RV 10 disconnected
Effect
Heating switched off
Solutions
Switch off the device
Clean the IR interface or remove the objects from around the IR interface
Switch on the device
E 3
Causes
Temperature inside the device is too high
Effect
Heating switched off
Solutions
Switch off the device
Let the device cool down
Switch on the device
E 9
Causes
Fault in storing the safety circuit set temperatures
Memory chip (EPROM) defective
Effect
Heating switched off
Solutions
Switch off the device
Let the device cool down
Switch on the device
E21
Causes
Safety relay does not open
Effect
Heating switched off
Solutions
Switch off the device and switch on again
E23
Causes
Adjustable safety circuit defective
Effect
Heating switched off
Solutions
Switch off the device
Let the device cool down
Switch on the device
E24
Causes
Set safety temperature exceeded
Effect
Heating switched off
Solutions
Switch off the device
Let the device cool down
Switch on the device
24
E25
Causes
The heater control circuit switch (TRIAC) is faulty�
The heater or the supply line is disconnected�
Effect
Heating switched off
Solutions
Switch off the device and switch on again
E26
Causes
Dry running
Effect
Heating switched off
Solutions
Switch off the device
Let the device cool down
Fill up the tempering medium
Switch on the device
E27
Causes
Error in calibration
Effect
Heating switched off
Solutions
Switch off the device and switch on again
E28
Causes
Breakage of the controller sensor
Effect
Heating switched off
Solutions
Switch off the device and switch on again
E29
Causes
Short-circuit in the safety sensor
Effect
Heating switched off
Solutions
Switch off the device and switch on again
E30
Causes
Short-circuit in the controller sensor
Effect
Heating switched off
Solutions
Switch off the device and switch on again
E31
Causes
Breakage of the safety sensor
Effect
Heating switched off
Solutions
Switch off the device and switch on again
E32
Causes
Temperature deviation is too big
Effect
Heating switched off
Solutions
Switch off the device and switch on again
If the actions described fails to resolve the fault or another error code is displayed then take one
of the following steps:
› contact the service department�
› send the device for repair, including a short description of the fault�
25

Advertisement

Table of Contents
loading