Error codes
The fault is shown by an error code on the display as following if the error occurs�
�
Proceed as follows in such cases:
› Turn off the device by using the main switch�
› Carry out corrective measures�
› Restart the device�
Error code | Causes | Effect | Solutions
E 2
Causes
›
In remote operation (PC), no communication between RV 10 and HB digital
›
IR connection to RV 10 disconnected
Effect
›
Heating switched off
Solutions
›
Switch off the device
›
Clean the IR interface or remove the objects from around the IR interface
›
Switch on the device
E 3
Causes
›
Temperature inside the device is too high
Effect
›
Heating switched off
Solutions
›
Switch off the device
›
Let the device cool down
›
Switch on the device
E 9
Causes
›
Fault in storing the safety circuit set temperatures
›
Memory chip (EPROM) defective
Effect
›
Heating switched off
Solutions
›
Switch off the device
›
Let the device cool down
›
Switch on the device
E21
Causes
›
Safety relay does not open
Effect
›
Heating switched off
Solutions
›
Switch off the device and switch on again
E23
Causes
›
Adjustable safety circuit defective
Effect
›
Heating switched off
Solutions
›
Switch off the device
›
Let the device cool down
›
Switch on the device
E24
Causes
›
Set safety temperature exceeded
Effect
›
Heating switched off
Solutions
›
Switch off the device
›
Let the device cool down
›
Switch on the device
24
E25
Causes
›
The heater control circuit switch (TRIAC) is faulty�
›
The heater or the supply line is disconnected�
Effect
›
Heating switched off
Solutions
›
Switch off the device and switch on again
E26
Causes
›
Dry running
Effect
›
Heating switched off
Solutions
›
Switch off the device
›
Let the device cool down
›
Fill up the tempering medium
›
Switch on the device
E27
Causes
›
Error in calibration
Effect
›
Heating switched off
Solutions
›
Switch off the device and switch on again
E28
Causes
›
Breakage of the controller sensor
Effect
›
Heating switched off
Solutions
›
Switch off the device and switch on again
E29
Causes
›
Short-circuit in the safety sensor
Effect
›
Heating switched off
Solutions
›
Switch off the device and switch on again
E30
Causes
›
Short-circuit in the controller sensor
Effect
›
Heating switched off
Solutions
›
Switch off the device and switch on again
E31
Causes
›
Breakage of the safety sensor
Effect
›
Heating switched off
Solutions
›
Switch off the device and switch on again
E32
Causes
›
Temperature deviation is too big
Effect
›
Heating switched off
Solutions
›
Switch off the device and switch on again
If the actions described fails to resolve the fault or another error code is displayed then take one
�
of the following steps:
› contact the service department�
› send the device for repair, including a short description of the fault�
25