Error Codes; Accessories; Maintenance - IKA C-MAG HS 4 Operating Instructions Manual

Hide thumbs Also See for C-MAG HS 4:
Table of Contents

Advertisement

3582825_C-MAG_072010:35 828 25 7/12/2010 4:03 PM 页 6

Error codes

Footstep during the enterprise a disturbance up, is indicated these
with the devices to HS and HP by an error message in the display
(C). Whenever an error message appears switch the device off to
cool down.
Error
Cause
code
E1
Inner temperature • switch off the unit and allow it
too high
E6
Motor seized
E9
Safety circuit
ETC
Have the device repaired if the error is not corrected using the
measures described or if another error code is displayed.

Accessories

Stirring organs
stirring bars:
Ø 6 mm, lengts to 15mm
Ø 8 mm, lengts to 50mm
Ø10mm, lengt to 80mm
Any other accessories
RS 1
Set of magnetic stirring bars
RSE
Stirring bar remover
H16V
Support rod
H16.1
Extension
H38
Holding rod
6
Solution
cool down
• Stirring bars ungeeignet
• Reduce the viscosity
• Foreign bodies in the equipment
remove
• Plug the contact plug (G)
• Plug the contact thermometer
• Connecting cable, plug or
contact thermometer defect,
exchange
H44
Cross sleeve
ETS-D5
Contact thermometer
Installing the support rod
The support rod is attached using the thraeded support bore (F).
• Screw nut M10 on to the support rod as far
as the stop
• Screw on the support rod as far as the
stop by hand
• Tighten the support rod and nut M10
using a flat wrench (SW17).
• Use bossheads to assemble
accessory parts or acces-
sory devices.

Maintenance

The appliance is maintenance-free.
Cleaning
Only use cleansing agents which have been recommended by IKA.
Use to remove:
Dyes
isopropyl alcohol
Construction materials water containing tenside/ isopropyl alcohol
Cosmetics
water containing tenside/ isopropyl alcohol
Foodstuffs
water containing tenside
Fuels
water containing tenside