Error Codes - IKA RV 10 control Operating Instructions Manual

Hide thumbs Also See for RV 10 control:
Table of Contents

Advertisement

Available languages

Available languages

1

Error codes

Any malfunctions during operation will be identified by an error message on the display.
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated. The lift can
be operated again.
Proceed as follows in such cases:
☞ Switch off device using the main switch
☞ Carry out corrective measures
☞ Restart device
Error
Communication Distillation stopped
error
No rotation
No temperature Distillation stopped
difference
Temperature
out of range
No vacuum
No venting
Water valve
not closed
Flow rate
out of range
Condenser
overload
Distillation
dry
Lift not in
end position
Calibration error
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short description of the fault.
042009
2
Effect
Distillation stopped,
no rotary drive
Distillation stopped
Distillation stopped
Vacuum cannot be controlled
Vacuum cannot be vented
Water runs continuously into cooling circuit
Distillation stopped
Distillation stopped
Distillate may be able to enter the vacuum
pump
Distillation stopped
Distillation stopped
Lift motor switches off
Calibration reset to factory setting
RV06.13
Seal FKM, PTFE-coated (1)
RV06.15
Seal PTFE (2)
PC 1.1
Cable
labworldsoft
Cause
IR interface obstructed
Heating bath switched off
Rotary drive does not start or
exhibits very large speed variation
No differential temperature increase of
cooling water
Differential temperature outside
measurement range (> 7 K)
Vacuum target value not reached
No venting of condenser
Water valve not closed
Flow rate of cooling water out of range
Condenser overload
The system detected that there was no
more medium in the evaporating flask
during the distillation.
Obstruction in path of lift
Out of calibration range (max. +/-0.5 K)
®
Correction
- Press ESC to confirm
- Switch on heating bath
- Check and clean IR interface
- Switch device off and on at mains switch
- Press ESC to confirm
- Check flow of cooling water
- Check that the cooling water hoses are fitted
correctly
- Press ESC to confirm
- Check that the cooling water hoses are fitted
correctly
- Press ESC to confirm
- Check that the glassware and hose connections
are properly sealed
- Switch on the pump
- Check the solenoid valve connections
- Switch off the device
- Vacuum sensor faulty
- Manually vent the glassware using the stop
valve
- Press ESC to confirm
- Check the connection
- Press ESC to confirm
- Readjust flow rate
- Switch off flow rate control
(for dry ice coolers)
- Press ESC to confirm
- Change process parameters, e.g.
- Increase pressure
- Reduce bath temperature
- Increase condenser flow rate within specified range
- Press ESC to confirm
- Alternatively, activate the "Drying" option
in the "Settings" menu
- Press ESC to confirm
- Check the path of the lift.
- Remove any obstructions.
- Press ESC to confirm
- Check that the sensor is connected correctly
- Carry out a new calibration procedure