Hepa Filter Pressure Reading Worksheet; Door Seal; Table 10. Hepa Filter Pressure Reading Worksheet - Despatch LCC Series Owner's Manual

Stackable oven
Hide thumbs Also See for LCC Series:
Table of Contents

Advertisement

LCC_LCD1-16/51 Oven Owner's Manual

7.3. HEPA Filter Pressure Reading Worksheet

A
B
Date Comments
Typical
Values
Filter first
installed

7.4. Door Seal

Check door seals are sitting properly and providing required sealing effect.
Check the gasket on the chamber door as follows.
1. Open the door.
2. Visually inspect the gasket sealing surface.
3. Squeeze the gasket between your fingers to check for damage and tears.
4. Close the door on a piece of paper at many points around the gasket. The
gasket should seal tight enough to require a slight pull to remove the
paper.
©2021 ITW EAE. All rights reserved. Despatch is a registered trademark in the U.S. and other
countries. No part of the contents of this manual may be reproduced, copied or transmitted in any form
or by any means without the written permission from ITW EAE - Despatch, unless for purchaser's
personal use.

Table 10. HEPA Filter Pressure Reading Worksheet.

C
D
Pressure
Pressure with 350
(inches
SCFH nitrogen purge
of water)
1.5-2.5" above value
2-3"
in Column C
C244, 320207, REV B
E
Pressure with 220
Oven
SCFH nitrogen
Temperature
maintain
0.3-0.8" above
60°C
value in Column C
F
69 of 82