LCC_LCD1-16/51 Oven Owner's Manual
Review the Protocol 3 Controller Owner's Manual for more information.
4.3. HEPA Filters
The Stackable LCC/LCD Ovens use HEPA (High Efficiency Particulate Air) filters to
limit particulate size in the work chamber to 0.3 microns or less. Two different
types of filters are typically used. While both filters are high-temperature filters in
that they are designed for use in thermal processing equipment, the "standard"
HEPA filter is rated to 260°C ( 500°F) while the "high-temperature" HEPA filter is
rated to 350°C (662°F).
LCC oven chamber temperature transitions must not exceed
1.5°C/minute to maintain class 100 chamber conditions. For
ramp rates greater than 1.5°C/minute and up to 5°C/minute, the
LCD model will maintain class 100 chamber conditions.
4.3.1. Definitions of Terms
• Binder: Organic substance used in filter construction to provide the media with
structural strength.
• Burn-Off: Process for eliminating the binder contained in the filter.
• D.O.P.: Dioctyl Phthalate - Aerosol particles of submicron size used in testing
phase to spot defects or measure filter efficiency.
• HEPA: High Efficiency Particulate Air
©2021 ITW EAE. All rights reserved. Despatch is a registered trademark in the U.S. and other
countries. No part of the contents of this manual may be reproduced, copied or transmitted in any form
or by any means without the written permission from ITW EAE - Despatch, unless for purchaser's
personal use.
Figure 5. Protocol 3 Operator Interface.
C244, 320207, REV B
27 of 82