6.6.2. MRC5000 Setup (Optional)

Refer to instructions provided recorder manufacturer for more
specific installation notes.
Temperature is retransmitted to the MRC5000 recorder from the controller. To
set up the recorder:
1. Ensure hardware jumper JU1 is in place for the 5 VDC setting (Refer to
MRC5000 Manual included).
2. Move MODE to PROG/TEST/CAL to display Prog.
3. Press ▼ twice to display Inps. Move to each Parameter Code using ▼or
▲. Adjust each Parameter Code using the settings in the table below.
After adjusting all settings, move MODE to RUN. Display on both the
4.
recorder and controller should read the same.
Parameter Code
Inps
Icor
diSP
dPOS
9
EUU
10
EUL
ChUP
ChLO
DFF
9
This value must match the setting Upper Retransmit Limit on the Protocol 3 Control page. For example, if Upper
Retransmit Limit is 752, EUU must read 752.
10
This value must match the setting Lower Retransmit Limit on the Protocol 3 Control page. For example, if Lower
Retransmit Limit is 32, EUL must read 32.
11
Change 0-400 chart paper to 0-800 chart paper. Depending on the equipment used, 0-600 paper may be used if
the maximum temperature is 500°F
64 of 82
©2021 ITW EAE. All rights reserved. Despatch is a registered trademark in the U.S. and other
countries. No part of the contents of this manual may be reproduced, copied or transmitted in any form
or by any means without the written permission from ITW EAE - Despatch, unless for purchaser's
personal use.

Table 8. MRC 5000 Settings.

Degrees C
17
0
On
0
400
0
400
0
1
LCC_LCD1-16/51 Oven Owner's Manual
Degrees F
17
0
On
0
752
32
11
800
0
1