14 Cleaning, Disinfection and Maintenance

14
Cleaning, Disinfection and Maintenance
14.1

Cleaning

62 / 85
The protective cover closes automatically. You can proceed with the
measurements.
During an examination: If you adjust the chin rest for a smaller head (the
head of a child, for example), the test eye may be in the way. Slide the
test eye to the side and then adjust the chin rest.
Cleaning and disinfection of the Pentacam® AXL is described in this chapter.
Sterilization is not required.
Heed the product descriptions and instruction manuals of products
and equipment you use to care for, clean, and disinfect the unit and/
or its accessories.
Do not clean the Pentacam® AXL with aggressive, chlorine
containing, abrasive or sharp cleaning agents.
Clean the cover glass from time to time to retain the high measurement
accuracy of the Pentacam® AXL and to avoid malfunctions. If the test
measurement shows a peak at 39 mm, a corresponding message will
occur, see
sec. 13, page
57.
Attention
Risk of electric shock if the Pentacam® AXL is not completely
disconnected from the mains for the cleaning.
Turn the Pentacam® AXL off,
Pull the power plug before cleaning. When disconnecting electrical
connections, pull on the respective plug and not on the cable itself.
Required materials:
Antistatic cleaning agent
Cleaner for painted surfaces: Mixture of equal parts of alcohol and
distilled water, possibly with a few drops of household detergent
Soft, lint-free cloth
Purified compressed air
commercial clean agent for the acrylic glass
Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04)