Guidance and Manufacturer's Declaration - Electromagnetic
Emisssions for the Pentacam® AXL
Guidance and manufacturer´s declaration electromagentic emissions
IEC 60601-1-2, 5.2.2.1, table 1
The OCULUS Pentacam® AXL is intended for operation in the electromagnetic environment specified below. The
user of the Pentacam® AXL should ensure that it is being used in such an environment.
Emissions test
RF emissions
CISPR 11
HF-emissions
CISPR 11
Harmonics emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
76 / 85
To be in compliance with the requirements of the IEC 60601-1-2. 6.1 and
6.2 the following types of equipment, accessories, power adapters and
cables must be used.
Order number
70100
05200320
05200210
(110 Volt)
05150150
70002
Compliance
Electromagnetic environment - guidance
Group 1
The Pentacam® AXL uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
Class B
Class A
complies
Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04)
Description
Pentacam® AXL
Cable with connector
plug, EU standard
Cable with connector
plug, US standard
Power adapter HMEG 49
Y cable with galvanic
isolation
2.5 m (98.4 in)
2.5 m (98.4 in)
24 V, 2,1A
2 m