Information for recording Scheimpflug images

Type of examination
Examination mode
Topography
3D scan
Pachymetry
3D scan
Analysis of the anterior
3D scan
chamber
Densitometry
3D scan
Axial length
AXL
30 / 85
If "Request 6x Alignment" is activated, the axial length measurement
starts only if the patient is fixating correctly during the complete scan.
After the Pentacam® AXL is aligned properly, the measurement starts
automatically.
"Request 6x Alignment" deactivated means: The axial length
measurements are performed without any interrupt.
"Request 6x Alignment" is activated by default and should just
deactivated if the patient has problems to fixate the red blinking light.
[Enh. Dyn.] checkbox in the "3D Scan" group box
Activating the "Enh. Dyn." function prolongs the exposure time per
Scheimpflug image. The advantage is an accurate representation of
phakic IOLs. If you select this recording mode, colours and
evaluations are neither calculated nor displayed.
"Fixation Target" slider
Use of the "Fixation Target" enables a better fixation of the patient.
For this, the active "Fixation Target" (8), i.e. the LED blinking red in
the middle of the blue slit, can be shifted in steps of 0.5 D. The object
is to offset defects in the patient's vision and ensure a simpler
method of fixation.
Images
Automatic measure-
ment
25-50
Yes
25-50
Yes
25-50
Yes
25-50
Yes
6
Yes
Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04)
Notes
Do not apply mydriatic drops.
Use the same number of images to en-
able a progress check, apply mydriatic
drops
Pay attention to eye status.