7 Setup and Connection

7
Setup and Connection
18 / 85
Attention
Risk of incorrect measurements/equipment damage due to improper
setup
Before first use, make sure the installation and connection of the
Pentacam® AXL are completed by our service or by a professional
authorized by OCULUS.
Note
Do not expose the Pentacam® AXL to any vibrations, shocks,
contaminants, moisture, or high temperatures.
Handle the optical device with care.
Set up the Pentacam® AXL so that the power plug is easy to access.
That way, you can easily disconnect it from the power supply for any
repairs or maintenance work.
Place the device so that direct light cannot affect the measurement.
Make sure the examination is free from light reflections. To achieve
this, darken the examination room.
Ambient temperature
The ambient conditions for operation are given in
Before installing the Pentacam® AXL, consider the transport, storage
and current room temperature, where the device is to be installed.
The difference between the temperature of the installation room and
the storage and transport temperature should not be more than 10°C
to avoid fogging of the internal optics.
If the difference in temperature is greater than 10° C, leave the
system at least for 6 hours until the temperature of the instrument
has adapted to room temperature.
Instruction Manual Pentacam® AXL (G/70100/EN 1121 Rev04)
sec. 19, page
72.