Three long beeps: Severe hardware error.
One short beep (1 second): Power up or Power down complete.
Four short beeps: There has been an error.
If you abort a BP measurement by pressing the Start/Stop button, you will hear one short beep. Then, after the
pressure has been completely dumped from the cuff, you will hear one long beep.
Can pulse oximetry, temperature measurement, a barcode scanner or printer be added later too?
Yes, you can add vital signs modules and accessories to the ADView 2 at any time.
Can EMR connectivity be established at a later date if not enabled as part of initial configuration?
Yes, EMR settings can be accessed via the Advanced Configuration Application at any time.
Why is no heart rate data showing on the display?
If the device is configured to use the Sp02 sensor as the pulse rate source, and the sensor is not attached to the
patient, no heart rate data will show on the display. Only dashes will appear, or MAP results if MAP measurement is
enabled. Note that MAP is not available in the United States.
How many measurements can be stored in the ADView 2 memory?
th
The ADView 2 can store 99 unique sets of measurements. After the 99
measurement has been saved, the next
measurement will have the Memory Location Identifier label "01" and will overwrite the most recent measurement
(measurement 99). After all results are cleared from the memory, the Memory Location Identifier also will return to
"01."

14. Technical Information

EMC Statement

This equipment needs special precautions regarding EMC and needs to be installed and put into service according to
the EMC information provided in this document. This equipment has been tested and found to comply with the limits
for medical devices to IEC60601-1-2: 2007. These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference
to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to other devices, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following
measures:
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from which the other device(s) are connected.
Consult the manufacturer or field service technician for help.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
WARNING: Use of accessories, transducers, and cables other than those specified may result in increased emissions
or decreased immunity of the ADView 2.
93-9005-00 rev 4
ADView 2 User Manual | 44
September 1, 2016