Temperature: 30.1 to 42.9 °C (86.2 – 109.2 °F)
SpO
: 70 – 100%
2
If measurement is above or below specified ranges:
Single Measurement Mode: The highest or lowest possible threshold range will blink in the appropriate BP
measurement field.
Averaging Mode:
will blink in addition to the status code "3" in the blood pressure measurement display.
Temperature: Depending on the out-of-range value, the display will indicate whether the value is "Hi" or "Lo."
SP02: Results falling below the measurement range will display "Lo" in the results field.

Service Centers

American Diagnostic Corporation
55 Commerce Drive
Hauppauge, NY 11788
Tel:
1.800.ADC.2670
1.631.273.9600
Fax:
1.631.273.9659
Email:
customerservice@adctoday.com
Web:
www.adctoday.com

13. Frequently Asked Questions

What is the expected battery life for the ADView 2?
When batteries are fully charged, the ADView 2 is designed to run on internal battery power for 200 measurement
cycles (BP, temperature and SpO
battery symbol is illuminated, there is between 11 percent and 40 percent battery power remaining.
Will the device automatically power down?
When no measurements have been taken or buttons pressed for one hour, the device will automatically power down
and enter a low-power state. This is considered "off."
What is the significance of different beeps I hear during operation?
One long beep (approximately 3 seconds): Successful end of BP measurement.
93-9005-00 rev 4
September 1, 2016
measurements) before requiring recharge. When only the bottom segment of the
2
ADView 2 User Manual | 43