System Description:

The ADView 2 spot check device with pulse oximetry module consists of a base ADView 2 device for the display of
SpO2 measurement data and other user information, either a ChipOx or Masimo SET SpO2 module which is attached
to the ADView 2 base unit as described in Section 2, and the specific reusable pulse oximeter probe and/or cable that
comes packaged with each module. Device setup is described in more detail in Section 2. Both the Masimo SET and
ChipOx SpO
modules are calibrated to display functional oxygen saturation (SpO
).
2
2
Taking SpO
Measurements:
2
1. Insert a finger (preferably the index, middle or ring finger) into the SpO2 sensor until the end of the finger reaches
the finger stop. Do not use the thumb.
2. The fingernail should face the side with the red light. Make sure that long fingernails do not interfere with proper
finger position.
An SpO
reading will be displayed after a few seconds. Remove the finger from the SpO
sensor, and the last
2
2
measurement will be displayed and flashed.
4. If the sensor is not removed from the finger, SpO
readings will be automatically stopped after 10 minutes of
2
continuous measurement, and the last measurement will be displayed and flashed.
5. To start a new measurement, re-apply the SpO
sensor to the finger as described in Step 1.
2
NOTE: If the signal quality from the sensor drops below 90%, no measurement will be displayed in the SpO
section
2
, and the percent symbol
will begin flashing. Check sensor for correct placement and usage.
Special Notes for Masimo SET SpO
Module:
2
NOTE: Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of one or
more of the patents relating to this device.
NOTE: If the accuracy of any measurement does not seem reasonable, first check the patient's vital signs by alternate
means and the check the MS board pulse oximeter for proper functioning.
NOTE: A functional tester cannot be used to assess the accuracy of a Masimo pulse oximeter sensor or pulse
oximeter. Using Clinical Dynamics SmartSat SpO2 simulator, the device can display the entire range of SpO2 values.
Masimo Sensors:
TM
Before use, carefully read the M-LNCS
DCI/DCIP sensor directions for use, which includes cleaning and disinfection
instructions.
T
Masimo M-LNCS
sensors are provided non-sterile.
Additional Masimo Information:
In addition to the General Principles of Operation mentioned above, the Masimo SET MS board pulse oximeter uses
the following principle:
The MS board decomposes the ratio of pulsatile absorbance (AC) to the mean absorbance (DC) at each of two
wavelengths, 660 nm and 905 nm (S(660) and S(905)) into an arterial signal plus a noise component and calculates
the ratio of the arterial signals without the noise:
S(660) = S1 + N1
S(905) = S2 + N2
R = S1/S2
93-9005-00 rev 4