613
Multiple patients have the same patient ID.
ADView 2 System:
300-302
Power off and repower on the ADView 2.
341-344
Power off and repower on the ADView 2.
Return for service.
Powering Device:
201
Run only from A/C power source.
Have battery service performed at earliest convenience.
202
Power off and repower on the ADView 2.
If repower attempt does not work, plug in power source and power off and repower on
ADView 2 again.
Retry reading. If issue persists, call Customer Service.
203-211
Clear the error and power off and repower on the ADView 2.
Replace battery.
212-214
Call Customer Service.
215
Charge battery.
Barcode Scanner:
701-706
Check and ensure proper barcode and patient ID was scanned
Rescan patient ID
Disconnect and reconnect cable from USB port on ADView 2.
Replace barcode scanner
Printing:
381
Close printer door.
382
Replace paper.
383- 387 & 388-390
Check cables. Reset printer. Replace printer.

Out of Range Measurements

The ADView 2 can measure BP that falls within the following ranges.
Systolic: Adult 40-260 mmHg, Pediatric 40-230 mmHg
Diastolic: Adult 20-200 mmHg, Pediatric 20-160 mmHg
Heart Rate: 30 - 220 BPM
The ADView 2 will display temperature and SpO
ranges.
93-9005-00 rev 4
September 1, 2016
within the following
2