Manual Mode
Example shows Manual Mode setup display. Three-digit number is the target cuff inflation pressure setting. The
default is 160 mmHg.
Select Mode
Press the Mode Toggle Button
you see the Manual Mode Icon
From single measurement mode, that will be two
button presses.

Step 4: How to Measure BP in Each Mode

When in the appropriate home screen (see above)
Single Measurement Mode
Ensure values are cleared from
display except for SYS value
representing current pressure in
cuff.
Press Start/Stop Button
.
Inflate/deflate cycle proceeds
automatically until results are
displayed.
A long beep will sound to indicate
measurement is complete.
93-9005-00 rev 4
September 1, 2016
Choose a Pressure Value
The pressure setting icon
until
flash. To keep the default maximum cuff inflation setting of
160 mmHG, press the Knob. Or choose another pressure level
.
between 100-280 mmHG and press the Knob to select. The
device will remember the pressure setting and display the
same pressure setting the next time SphygMode is used.
Averaging Measurement Mode
When selecting this mode the
will flash to allow the user to decide
how many reading they would like
averaged.
Select number of readings by
rotating the Selection Knob and
then pushing the knob in.
The minute number will flash and
the user can select how many
minutes until the first reading starts
by rotating the Selection Knob and
pushing in.
The user can then select the time
between readings by rotating the
Selection Knob and pushing in.
The timer will then count down the
seconds until the first reading.
After all measurements are taken
and averaged, results will display.
A long beep will sound to indicate
measurement is complete.
and Selection Knob will
Manual Mode
Put stethoscope over artery.
Press Start/Stop
inflate cuff automatically. Similar to
an analog sphyg gauge, the display
shows current pressure in cuff.
When cuff reaches max pressure,
deflation will begin automatically.
Listen for K-sounds while watching
displayed pressure. Note SYS and
DIA.
Press Start/Stop
quickly dump remaining pressure
from the cuff.
ADView 2 User Manual | 24