Cleaning And Disinfection; Sensors And Cables; Sensor Holders - KaVo ProXam iS Instructions For Use Manual

Table of Contents

Advertisement

10 Cleaning and disinfection

Recommended disinfectants

10.1 Sensors and cables

10.2 Sensor holders

Instructions for use
NOTE
Before cleaning the system, always check that the X-ray unit and the digital
sensor system are off (the control box indicator light is off).
For disinfection manufacturer recommends the following disinfectants for
wiping:
CaviWipes (Metrex Research, USA)
Dürr FD 322 (Dürr Dental AG, Orochemie GmbH)
Dürr FD 333 (Dürr Dental AG, Orochemie GmbH
Cidex Opa (Johnson & Johnson)
All recommended disinfecting agents have been tested and found to be
harmless to the surfaces.
The sensors allow enhanced infection control in surgery.
NOTE
Always use appropriate instruments for cleaning the sensors.
NOTE
It is mandatory to carefully follow the cleaning and disinfection
recommendations for not to damage the sensors.
CAUTION
The sensors cannot be sterilized in autoclave or UV oven.
Wipe the sensor surface with a soft cloth damped into a disinfectant solution.
The recommended disinfectant solutions are Dürr System Hygiene FD 322
or FD 333 or similar product.
If more effective disinfection or cold sterilization is preferred for cleaning, we
recommend the Johnson & Johnson Cidex Opa high level disinfectant.
NOTE
Follow carefully the manufacturer's recommendations on disinfectant liquids.
Use a new disposable hygienic sheath for every sensor usage.
NOTE
The sensor connector can be cleaned using a soft cloth.
For cleaning the sensor holders refer to the manual supplied with the sensor
holder package.
10 Cleaning and disinfection
KaVo ProXam iS
15

Advertisement

Table of Contents
loading

Table of Contents