Presys DMY-2030-F Technical Manual page 27

Frequency digital indicator
Table of Contents

Advertisement

PRESYS
⏐ Instruments
RANGE - selects the range of retransmission output as 20mA, 5V or 10V. The
relation between the engineering unit and electric signal generated in the terminals is
defined in the same way as shown for the input scale configuration. Define two points P1
(Eng Low, Lim Low) and P2 (Eng High, Lim High) as it is illustrated in Figure 15.
Eng Low is the indication in the display in engineering units associated with the electric
signal Lim Low, and Eng High is the indication in display in engineering units associated
with the electric signal Lim High. Note that Lim Low and Lim High are defined in
percentage of output range and that the output signal saturates in these points.
(*) % FULL-SCALE OUTPUT SIGNAL
Level 5 - Calibration
Level 5 is described on section 4.5 on Calibration.
Level 6 - RS
Refer to the communication manual.
Operation
Page 24
Fig. 15 - Analog output configuration
DMY-2030-F / DMY-2030-F Light