Trouble-shooting
Error
Display
no.
Below lower alarm threshold for
E154
period exceeding alarm delay
Above upper alarm threshold for
E155
period exceeding alarm delay
Actual value undershoots alarm
E156
threshold for longer than the set
permissible maximum period
Actual value exceeds alarm
E157
threshold for longer than the set
permissible maximum period
E162
Dosage stop
Dynamic range of pH converter
E164
exceeded
Dynamic range of reference
E166
converter exceeded
Warning: ISFET leakage
E168
current > 200 nA
Flow in main stream too low or
E171
zero
Switch-off limit for current input
E172
exceeded
E173
Current input < 4 mA
E174
Current input > 20 mA
E175
SCS glass warning
E177
SCS reference electrode warning
1)
If this error occurs, there is no possibility of starting a cleaning session (field F8 not applicable with this error).
80
Tests and/or remedial measures
Perform manual comparison measurement if
necessary. Service sensor and recalibrate.
Check settings in the CURRENT INPUT and
CHECK function groups.
Check cable and sensor.
Check cable and sensor.
Check ISFET for abrasion and air-tightness,
replace as soon as possible.
Restore flow.
Check process variables at sending
measuring instrument. Change range
assignment if necessary.
Check process variables at sending
measuring instrument.
Check process variables at sending
measuring instrument. Change range
assignment if necessary.
Check electrode for glass breakage or hair-
line cracks;
Check medium temperature.
Measurement can continue until the error
occurs.
Check electrode for contamination and
damage; clean electrode; measuring can
continue until the error occurs.
Alarm contact
Error current
Facty
User
Facty
User
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
No
No
Yes
No
Yes
No
Yes
No
Yes
No
No
No
No
No
OPM223/253
Autom.
cleaning start
Facty
User
No
No
No
No
No
No
No
No
No
No
No
No