OEM OPM223 Operating Instructions Manual page 79

Transmitter for ph and orp
Table of Contents

Advertisement

OPM223/253
Error
Display
no.
Calculation of calibration
E041
parameters aborted
Distance of calibration value for
E042
buffer pH 2 from zero (pH 7) too
short
Distance between calibration
E043
values for pH 1 and pH 2 too
short
Stability requirement for
E044
calibration not fulfilled
E045
Calibration aborted
Below main parameter
E055
measuring range
Main parameter measuring
E057
range exceeded
Below temperature measuring
E059
range
Temperature measuring range
E061
exceeded
E063
Below current output range 1
E064
Current output range 1 exceeded
E065
Below current output range 2
E066
Current output range 2 exceeded
E067
Set point exceeded controller 1
E068
Set point exceeded controller 2
E069
Set point exceeded controller 3
E070
Set point exceeded controller 4
E080
Current output 1 range too small
E081
Current output 2 range too small
E085
Incorrect setting for error current
E100
Current simulation active
E101
Service function yes
E102
Manual mode active
E106
Download yes
E116
Download error
E147
Sensor communication faulty
E152
PCS alarm
Tests and/or remedial measures
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device
and measuring cable with simulator.
Use buffer solution that is at least Δ pH = 2
from electrode zero point.
Use buffer solutions that are at least Δ pH = 2
apart.
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device
and measuring cable with simulator.
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device
and measuring cable with simulator.
Check measurement and connections; check
device and measuring cable with simulator if
necessary.
Check configuration in the "Current outputs"
menu; check measurement and connections;
check device and measuring cable with
simulator if necessary.
Check configuration.
Increase range in "Current outputs" menu.
If the current range "0 ... 20 mA" was selected
in field O311, the error current "2.4 mA" may
not be set.
Switch off service function or switch device off
and then on again.
Wait for download to finish.
Repeat download.
Check that the sensor is correctly connected,
the cable ends are correctly wired at the
terminals and the cable is not damaged.
Check sensor and connection.
Trouble-shooting
Alarm contact
Error current
Facty
User
Facty
User
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
No
No
No
No
?
?
No
No
No
No
No
No
No
No
No
No
No
No
No
No
Autom.
cleaning start
Facty
User
1
1
1
1
1
No
No
No
No
No
No
No
No
No
No
No
No
1
1
?
?
1
1
1
1
1
No
No
79