Commissioning
User input
13. In A5, select the temperature sensor the
electrode used has, e.g. "Pt 100" for a glass
electrode.
Press
The display returns to the initial display of the
"Setup 1" function group.
14. Press
15. Press
16. In B1, select the type of temperature
compensation for the process, e.g. ATC for
automatic temperature compensation.
Press
If you have chosen ATC, the menu jumps
automatically to field B3.
17. In B3, select the type of temperature
compensation for the calibration, e.g. ATC for
automatic temperature compensation.
Press
18. The current temperature is displayed in B4. If
necessary, calibrate the temperature sensor to
an external measurement.
Press
19. The difference between the measured and
entered temperature is displayed.
Press
The display returns to the initial display of the
"Setup 2" function group.
20. Press
measurement mode.
68
F
to confirm your entries.
S
to get to the "Setup 2" function group.
F
to make your settings for "Setup 2".
F
to confirm.
F
to confirm.
F
to confirm.
F
.
P
simultaneously to switch to the
Setting range
Display
(Factory settings,
bold)
Pt 100
Pt 1K
NTC 30K
None
SETUP 2
ATC
MTC
°C-Proces
ATC
MTC
°C-Cal
Actual value
displayed and
entered
RealTemp
-50.0 ... 150.0 C
0.0 C
-5.0 ... 5.0 C
Temp.Offs
OPM223/253
Pt100
A5
Temp
B
ATC
B1
ATC
B3
°C
25.0
B4
°C
0.0
B5