OPM223/253
Coding
Field
Enter set point 1 (or
R262
2)
Enter relay
R263
switching capacity
K
1 (or K
2)
p
p
Enter integral action
time T
1 (or T
n
R264
(0.0 = no
I-component)
Enter derivative
action time T
R265
T
2)
v
(0.0 = no
D-component)
Select pulse length
R266
or pulse frequency
R267
Enter pulse interval
Enter maximum
R268
pulse frequency of
the adjuster
Enter minimum
R269
switch-on time t
R2610 Enter process type
Setting range
(Factory settings, bold)
pH 6.00
pH -2.00 ... 16.00
1.00
0.10 ... 20.00
2)
0.0 min
n
0.0 ... 999.9 min
1 (or
v
0.0 min
0.0 ... 999.9 min
Len = pulse length
Freq = pulse frequency
Curr = current output 2
10.0 s
0.5 ... 999.9 s
-1
120 min
-1
60 ... 180 min
0.3 s
0.1 ... 5.0 s
ON
Batch
Inlne
Display
pH
6.00
R262
Setpoint1
1.00
R263
Kp Rel1
0.0
min
R264
Tn Rel1
0.0
min
R265
TV Rel1
len
R266
Oper.Mode
s
10.0
R267
Puls Per.
1/min
120
R268
Max.PFreq
s
0.3
R269
Min.PTime
Batch
%
R2610
Proc.Type
Operation
Info
Relay assignment 1 and 2 for neutra
controller:
Rel1 = set point 1
Rel2 = set point 2
Relay assignment 1 and 2 for neutra
controller:
Rel1 = Kp1
Rel2 = Kp2
Relay assignment 1 and 2 for neutra
controller:
Rel1 = Tn1
Rel2 = Tn2
Relay assignment 1 and 2 for neutra
controller:
Rel1 = Tv1
Rel2 = Tv2
Pulse length e.g. for solenoid valve,
pulse frequency e.g. for solenoid dosing
pump, see "Actuating signal outputs"
section.
Curr = current output 2 can only be
selected if O2 = Contr.
This field only appears if pulse length is
selected in R266.
If pulse frequency is selected, R267 is
skipped and entries continue with R268.
This field only appears if pulse
frequency is selected in R266.
If pulse length is selected, R268 is
skipped and entries continue with R269.
This field only appears if pulse length is
selected in R266.
Batch = discontinuous process
Inlne = continuous process
There is no further dosing in the setting
range in batch mode. The I-component
is decreased.
Dosing continues in the setting range in
inline mode. The I-component is
effective.
55