Quick Start-Up - OEM OPM223 Operating Instructions Manual

Transmitter for ph and orp
Table of Contents

Advertisement

OPM223/253
6.4
After power-up, you must make some settings to configure the most important functions of the
transmitter which are required for correct measurement. The following section gives an
example of this.
User input
1. Press the
2. Enter the code 22 to edit the setup. Press
3. Press
group.
4. Press
5. In S1, select your language, e.g. "ENG" for
English.
Press
6. Press
function group.
7. Press
group.
8. Press
"Setup 1".
9. In A1, select the desired mode of operation, e.g.
"pH".
Press
10. In A2, select the type of connection for your
sensor. See the "Sensor connection" section for
this.
Press
11. In A3, enter the damping factor.
Measured value damping averages the
individual measured values and serves to
stabilise the display and the signal output. Enter
"1" if no measured value damping is required.
Press
12. In A4, specify the type of sensor that you are
using, e.g. "Glass" for glass electrode.
Press

Quick start-up

F
key.
S
until you get to the "Service" function
F
to be able to make your settings.
F
to confirm.
P
simultaneously to exit the "Service"
S
until you get to the "Setup 1" function
F
to be able to make your settings for
F
to confirm.
F
to confirm.
F
to confirm.
F
to confirm.
Setting range
Display
(Factory settings,
bold)
F
.
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
pH
ORP mV
ORP %
sym = symmetrical
asym = asymmetrical
1
1 ... 60
Glass
ISFET
Commissioning
S
SERVICE
ENG
S1
Language
A
SETUP 1
pH
A1
Oper.Mode
sym
A2
Wiring
1
A3
Damping
Glass
A4
Sensor
67

Advertisement

Table of Contents
loading

This manual is also suitable for:

Opm253

Table of Contents