OEM OPM223 Operating Instructions Manual page 57

Transmitter for ph and orp
Table of Contents

Advertisement

OPM223/253
Coding
Field
Reset the device to
the basic settings
S9
S10
Perform device test
Reference voltage
S11
is displayed
Select AC
S12
frequency
5.4.8
Coding
Field
SERVICE 2
E
function group
E1
Select module
E111
E121
Software version is
E131
displayed
E141
E151
E112
E122
Hardware version is
E132
displayed
E142
E152
Setting range
(Factory settings, bold)
No
Sens = sensor data
Facty = factory settings
No
Displ = display test
Current value in mV
50 Hz
60 Hz
Service 2
Setting range
(Factory settings, bold)
Contr = controller (1)
Trans = transmitter (2)
Main = power unit (3)
Rel = relay module (4)
Display
no
S9
S.Default
no
S10
Test
25
mV
S11
uncompRef
50
Hz
S12
AC Freq
Display
E
SERVICE 2
Contr
E1
Select
xx.xx
E111
SW-Vers.
xx.xx
E112
HW-Vers.
Operation
Info
Sens = last calibration is deleted and is
reset to factory setting.
Facty = all data (apart from A1 a. S1) are
deleted and reset to the factory setting!
This is used to check the reference
potential. Value > 50 mV indicates
galvanic voltage in the medium.
High values (> 1000 mV) may falsify the
measured value.
Only select 60 Hz if the frequency of the
voltage at the place of use is 60 Hz, the
measured value is fluctuating or sporadic
SCS errors are reported.
Note
Information on the device version
If E1 = contr: instrument software
If E1 = trans, main, rel: module firmware
Only display function
57