OEM OPM223 Operating Instructions Manual page 54

Transmitter for ph and orp
Table of Contents

Advertisement

Operation
Coding
Field
R253
Enter pre-rinse time
R254
Enter cleaning time
Enter post rinse
R255
time
Enter number of
R256
repeat cycles
R257
Enter pause time
Enter minimum
R258
pause time
Enter number of
cleaning cycles
R259
without cleaning
agent (economy
function)
Configure
R2 (6)
neutralisation
controller
Switch function of
R261
R2 (6) off or on
54
Setting range
Display
(Factory settings, bold)
20 s
0 ... 999 s
10 s
0 ... 999 s
20 s
0 ... 999 s
0
0 ... 5
360 min
1 ... 7200 min
120 min
1 ... R257 min
0
0 ... 9
LC PV = limit contactor pH/
ORP (1)
LC C = limit contactor T (2)
PID controller (3)
Timer (4)
Clean = Chemoclean (5)
Neutra controller (6)
Off
On
Info
s
20
Rinsing with water takes place.
R253
PreRinse
10
s
Cleaning with cleaning agent and water
R254
takes place.
CleanTime
20
s
Rinsing with water takes place.
R255
PostRinse
0
R253 ... R255 is repeated.
R256
Rep.Rate
The pause time is the time between two
360
min
cleaning cycles (see "Timer function"
R257
section).
PauseTime
The minimum pause time prevents
min
120
constant cleaning if an external cleaning
R258
start is present.
Min.Pause
After cleaning with cleaner, up to 9
0
cleaning sessions can be carried out
R259
with water only until the next cleaning
EconomyCl
session with cleaner takes place.
Only for A1 = pH.
If neutra controller is selected for Rel1,
only neutra controller is offered for Rel2.
Neutr
By confirming with ENTER, another
R2
relay function already switched on is
Sel.Type
switched off and its settings are reset to
the factory settings.
off
R261
Function
OPM223/253