Wiring
"
Caution!
• Terminals marked NC may not be wired.
• Unmarked terminals may not be wired.
!
Note!
Please label the sensor terminal block with the sticker provided.
4.1.2
You require screened special measuring cables to connect pH and ORP electrodes to the
transmitter. The following multi-core and ready-to-use cable types can be used:
Sensor type
Electrode without temperature sensor
Electrode with temperature sensor Pt 100
and TOP 68 plug-in head
ISFET sensor with temperature sensor
Pt 100 / Pt 1000 and TOP 68 plug-in head
pH individual electrode with separate
reference electrode and separate
temperature sensor
Structure and termination of the measuring cables
OPK1
OPK2
OPK9
OPK12
Fig. 12:
!
Note!
For further information on the cables and junction boxes, please refer to the "Accessories"
section.
18
Measuring cable and sensor connection
PM
pH / Redox
PM
pH or Redox
Ref. or Redox
Pt 100
PM
pH + Pt 100 / Pt 1000
PM
pH + Pt 100 / Pt 1000
Structure of the special measuring cables
Cable
Extension
OPK1
VBA / VBM box + OYK71 cable
OPK9
VBA / VBM box + OYK71 cable
OPK12
VBA / VBM box + OYK12 cable
OPK2
VBA / VBM box + PMK cable
OPM223/253
Coax Ref.
pH or Redox
BN PM
BN PM
pH or Redox
RD Redox or Ref.
BU Ref.
GN Temp.
YE
Temp.
WH Temp.
Coax BK
Ref.
inner pH
GN Temp.
WH Temp.
YE
Temp.
BN
PM
Coax BK
Ref
inner
SRC IsFET
GN Temp.
WH Temp.
YE
Temp.
RD
DRN
BN
PM
C07-OPM2x3xx-04-06-00-en-004.eps

Advertisement

Table of Contents
loading

This manual is also suitable for:

Opm253