OEM OPM223 Operating Instructions Manual page 50

Transmitter for ph and orp
Table of Contents

Advertisement

Operation
Coding
Field
Switch function of
R211
R2 (1) off or on
Enter the switch-on
R212
point of the contact
Enter the switch-off
R213
point of the contact
R214
Enter pick-up delay
Enter drop-out
R215
delay
Enter alarm
R216
threshold
Display status for
R217
limit contactor
Configure limit
contactor for
R2 (2)
temperature
measurement
Switch function of
R221
R2 (2) off or on
50
Setting range
Display
(Factory settings, bold)
Off
On
pH 16.00
pH -2.00 ... 16.00
1500 mV
-1500 ... 1500 mV
100.0 %
0.0 ... 100.0 %
pH 16.00
pH -2.00 ... 16.00
1500 mV
-1500 mV ... 1500 mV
100.0 %
0.0 ... 100.0 %
0 s
0 ... 2000 s
0 s
0 ... 2000 s
pH 16.00
pH -2.00... 16.00
1500 mV
-1500 ... 1500 mV
100.0 %
0.0 ... 100.0 %
MAX
MIN
LC PV = limit contactor pH/
ORP (1)
LC C = limit contactor T
(2)
PID controller (3)
Timer (4)
Clean = Chemoclean (5)
Neutra controller
Off
On
Info
off
All the settings are retained.
R211
Function
Never set the switch-on point and the
pH
16.00
switch-off point to the same value!
R212
(Only the operating mode selected in A1
On value
is displayed.)
Entering a switch-off point selects either
a Max contact (switch-off point < switch-
on point) or a Min contact (switch-off
pH
16.00
point > switch-on point), thereby
R213
implementing a hysteresis that is
Off value
constantly required (see "Illustration of
the alarm and limit functions" figure).
s
0
R214
On Delay
s
0
R215
Off Delay
If the alarm threshold is undershot/
overshot, this triggers an alarm with the
pH
16.00
error message and error current at the
R216
transmitter (note alarm delay in field F3).
A.Thresh
If defined as a Min contact, the alarm
threshold must be < switch-off point.
MAX
Display only.
R217
LC State
By confirming with ENTER, another
LC °C
relay function already switched on is
R2
switched off and its settings are reset to
Sel.Type
the factory settings.
off
R221
Function
OPM223/253

Advertisement

Table of Contents
loading

This manual is also suitable for:

Opm253

Table of Contents