OEM OPM223 Operating Instructions Manual page 36

Transmitter for ph and orp
Table of Contents

Advertisement

Operation
Value pair
1
2
3
4
5
6
7
8
9
Coding
Field
CURRENT
O
OUTPUT function
group
Select current
O1
output
Select measured
O2
variable for 2nd
current output
Enter or output
O3
O3 (1)
linear characteristic
Select current
O311
range
36
Current output 1
pH / mV /
Current
% / C
[mA]
Setting range
(Factory settings, bold)
Out1
Out 2
°C
pH
mV
Contr
Lin = linear (1)
Sim = simulation (2)
Tab = table (3)
4 ... 20 mA
0 ... 20 mA
Current output 2
Distance
pH / mV /
per mA
% / C
Display
O
OUTPUT
Out1
O1
Sel.Out
°C
O2
Parameter
lin
O3
Sel.Type
4-20
O311
Sel.Range
OPM223/253
Current
Distance
[mA]
per mA
Info
Configuration of the current output.
A characteristic can be selected for
every output.
R237/R 266 = curr (current output 2)
can only be selected if O2 = Contr is
selected (relay board required).
The characteristic can have a positive
or negative slope for the measured
value output.
In the case of actuating variable output
(O2 = Contr), an increasing current
corresponds to an increasing actuating
variable.