Operation
Coding
Field
Enter control gain
R233
K
p
Enter integral action
time T
n
R234
(0.0 = no I-
component)
Enter derivative
action time T
R235
(0.0 = no D-
component)
Select controller
R236
characteristic
Select pulse length
R237
or pulse frequency
R238
Enter pulse interval
Enter maximum
R239
pulse frequency of
the adjuster
Enter minimum
R2310
switch-on time t
R2311 Enter basic load
52
Setting range
(Factory settings, bold)
1.00
0.01 ... 20.00
0.0 min
0.0 ... 999.9 min
0.0 min
v
0.0 ... 999.9 min
dir = direct
Inv = inverse
len = pulse length
Freq = pulse frequency
Curr = current output 2
10.0 s
0.5 ... 999.9 s
-1
120 min
-1
60 ... 180 min
0.3 s
0.1 ... 5.0 s
ON
0 %
0 ... 40 %
Display
1.00
R233
Kp
min
0.0
R234
Time Tn
min
0.0
R235
Time Tv
dir
R236
Direction
len
R237
Oper.Mode
s
10.0
R238
PulsePer.
1/min
120
R239
Max.PFreq
s
0.3
R2310
Min.PTime
0
%
R2311
BasicLoad
OPM223/253
Info
See "P(ID) controller" section.
See "P(ID) controller" section.
With every Hold, the I-component is set
to zero. Although Hold can be
deactivated in field S2, this does not
apply for Chemoclean and timer!
See "P(ID) controller" section.
The setting is required depending on the
control deviation (upward or downward
deviation, see "Chemoclean function"
section).
Pulse length e.g. for solenoid valve,
pulse frequency e.g. for solenoid dosing
pump, see "Actuating signal outputs"
section.
Curr = current output 2 can only be
selected if O2 = Contr.
This field only appears if pulse length is
selected in R237.
If pulse frequency is selected, R238 is
skipped and entries continue with R239.
This field only appears if pulse
frequency is selected in R237. If pulse
length is selected, R239 is skipped and
entries continue with R2310.
This field only appears if pulse length is
selected in R237.
When you select the basic load, you
enter the desired dosing quantity.
100% basic load corresponds to:
– Constantly on for R237 = len
– Fmax at R237 = feq
– 20 mA at R237 = curr