Trouble-Shooting; Trouble-Shooting Instructions; System Error Messages - OEM OPM223 Operating Instructions Manual

Transmitter for ph and orp
Table of Contents

Advertisement

Trouble-shooting

9
9.1
The transmitter constantly monitors its functions itself. If an error occurs which the device
recognises, this is indicated on the display. The error number is under the unit display of the
main measured value. If more than one error occurs, you can call these up with the MINUS key.
Refer to the "System error messages" table for the possible error numbers and remedial
measures.
Should a malfunction occur without any transmitter error message, please refer to the
"Process-specific errors" or the "Device-specific errors" tables to localise and rectify the error.
These tables provide you with additional information on any spare parts required.
9.2
You can display and select the error messages with the MINUS key.
Error
Display
no.
E001
EEPROM memory error
Instrument not calibrated,
calibration data invalid, no user
E002
data invalid (EEPROM error),
instrument software not suitable
to hardware (controller)
E003
Download error
Instrument software version not
E004
compatible with module
hardware version
Transmitter malfunction,
instrument software not
E007
compatible with transmitter
version
SCS alarm:
E008
Glass electrode: glass breakage
ISFET: leakage current > 400 nA
Temperature sensor defective,
E010
not connected or short-circuited
E030
SCS reference electrode warning
Slope range exceeded or below
E032
range
E033
pH value zero too low or too high
ORP offset range exceeded or
E034
below range
78
Trouble-shooting

Trouble-shooting instructions

System error messages

Tests and/or remedial measures
1. Switch device off and then on again.
2. Load device software compatible with the
hardware (with optoscope, see
"Optoscope service tool" section).
3. Load measurement-parameter specific
device software.
4. If the error persists, send in the device for
repair to your local supplier or replace the
device.
Invalid configuration. Repeat download, check
optoscope.
Load software compatible with hardware
Load measurement-parameter specific device
software.
Check glass electrode for glass breakage or
hair-line cracks;
Inspect electrode plug-in head for moisture
and dry if necessary;
Check medium temperature.
Replace ISFET.
Check temperature sensor and connections;
check device and measuring cable with
temperature simulator if necessary.
Check correct option selected in field A5.
Check reference electrode for contamination
and damage; clean electrode.
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device
and measuring cable with simulator.
Alarm contact
Error current
Facty
User
Facty
User
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
No
No
Yes
No
Yes
No
Yes
No
OPM223/253
Autom.
cleaning start
Facty
User
1
1
No
No
1
No
No
No
1
1
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Opm253

Table of Contents