4.Sequence Function
 Step transition method
 Selecting the trigger source
 Signal output
PLZ-5W
The available load setting transition methods are Ramp and Immediate. With Immediate,
the load setting transitions step-wise from the setting of the previous step. With Ramp, the
load setting transitions in a ramp from the setting of the previous step. For step 1, the load
setting start point is 0 A, 0 S, 0V, or 0 W.
Example: Level: 10A, Dwell: 100 ms
Immediate
10 A
100 ms
For Wait(pre) (pre-trigger) and Wait(post) (post-trigger), select the trigger source. If you
set the Wait(pre) trigger source, the program pauses before executing the step and waits
for a trigger. If you set the Wait(post) trigger source, the program pauses after executing
the step and waits for a trigger. If you do not set anything, the transition method is set to
Immediate, and the execution of the step starts without the application of a trigger.
Item
Source
Value
Wait(pre)
MSYNc
Synchronizes the resuming of sequences on synchronized PLZ-5Ws
(p.53).
Executes the step at the timing at which a signal is input to pin 13 of the
1
Digital2
EXT CONT connector.
2
Executes the step when the *TRG command is sent from a PC, etc.
BUS
Wait(post)
Trig IN
Executes the next step when a trigger is received (p.78).
1
Only when Direction of Digital2 is set to Input (p.87).
2
If set to BUS, other trigger sources cannot be set.
For Generate and →, set the signal output during the step execution.
Item
Source
Generate
TA Link
Trig Out
Digital0
Digital1
1
Digital2
1
Only when Direction of Digital2 is set to Input
For Trig Out, the trigger is output when a step is executed.
For Digital0, Digital1, Digital2, the signal is set to high level during a step.
Example: When Trig Out, Digital0, Digital1, and Digital2 are set in step 2.
Load setting
Set
0
Step 1
TRIG OUT
DIGITAL 0
DIGITAL 1
DIGITAL 2
Ramp
10 A
100 ms
Value
Output TA Link triggers. If source is set to TA Link on (p.37),
measurement starts at the step execution timing.
Trigger output from the TRIG OUT terminal.
The pin 11 signal of the EXT CONT connector is set to high level.
The pin 12 signal of the EXT CONT connector is set to high level.
The pin 13 signal of the EXT CONT connector is set to high level.
(p.87)
Set: Load setting based on step setting
Time
Step 2
Step 3
Setting Steps