Kikusui PLZ-5W Series User Manual page 114

Electronic load
Hide thumbs Also See for PLZ-5W Series:
Table of Contents

Advertisement

Operating Area
114
Transition of the operating point: Overcurrent protection (OCP)
operation
Operating area
A
A
A
1
2
V
1
R
1
R
2
R
F
Logarithmic scale
Input current (A)
If the OCP setting I
OCP
resistance is decreased (R
the voltage of the constant-voltage power supply at V1, the operating point moves along seg-
ment AF (A
→A
→F). When point F is reached, OCP is activated.
1
2
At this point, two types of operation are available on the PLZ-5W depending on which action
is set to be executed at OCP detection.
If the action set to be executed at OCP detection is Trip, load off is executed.
If the action set to be executed at OCP detection is Limit, the PLZ-5W sinks current as a con-
stant current load at point F. Even if you attempt to increase the input current by decreasing
the resistance, the current is limited at point F. If you decrease the input current by increasing
the resistance, the OCP is cleared. The PLZ-5W returns to CR mode, and the operating point
moves along segment AF.
Trip
Turns off the load (stops the current flow). The PLZ-5W no longer operates as a load.
Point F
CR mode ends. OCP continues, and the PLZ-5W sinks current as a constant current
Limit
load.
Constant current line
(OCP setting)
OCP trip point
F
B
Constant resistance line
(Constant resistance settings: R
I
I
OCP
B
is smaller than the current value I
→R
→R
) and the input current (load current) is increased with
1
2
F
10.Appendix
, R
, R
)
1
2
F
at point B, when the PLZ-5W
B
PLZ-5W

Advertisement

Table of Contents
loading

This manual is also suitable for:

Plz205wPlz405wPlz1205wPlz2405wb

Table of Contents