Switching Function
30
Setting the switching interval with the duty ratio
Set the frequency (or the time of one cycle) and duty ratio.
1
Press SOURCE, Switching, and then Freq.
2
Use the numeric keypad or the rotary knob to enter the frequency (1 Hz
to 100 kHz).
The input switches between frequency [Hz] input and 1 cycle's time [s] input every
time you press Freq.
3
Press Duty|Th.
4
Use the numeric keypad or the rotary knob to enter the duty ratio (1 % to
99 %).
The minimum switching interval is 5 μs.
This sets the switching interval.
Setting the switching interval with the HIGH side operation times
Set the frequency (or the time of one cycle) and the operation time on the HIGH side.
1
Press SOURCE, Switching, and then Freq twice.
2
Use the numeric keypad or the rotary knob to enter the frequency (1 Hz
to 100 kHz).