Measurement Function
38
 Trigger processing procedure
The general trigger processing procedure is illustrated in the following figures.
Abort
Power ON
Press Initiate
yes
count=0
Immediate
BUS
DIGITAL2
Source
Event detector
TALink
MSync
timer = 0
Acquire action
(Sense Aperture)
Timer start
 Trigger application examples
Example: Immediate is set for Source, Interval to Disable, and measurement value acqui-
sition is repeated three times.
Initiate key press
Measurement
value acquisition
Count = 1
Delay
Sense Aperture
Example: BUS/DIGITAL2/TALink/MSync is set for Source, Interval to Disable, measure-
ment result acquisition is repeated three times.
Initiate key press
Trigger application
Delay
Example: Source is set to Immediate, Interval to Enable, and Count to 3.
Initiate key press
Count = 1
Delay
Sense Aperture
Interval Time
Delay
yes
timer=
Interval Time?
no
count = count+1
Measurement
Measurement
value acquisition
value acquisition
Count = 2
Sense Aperture
Sense Aperture
Trigger application
Measurement
Measurement
value acquisition
value acquisition
Count = 1
Count = 2
Sense Aperture
Delay
Sense Aperture
Count = 2
Sense Aperture
Interval Time
2.Basic Functions
Other
Immediate
Trigger source
Disable
Enable
Interval
no
yes
count =
Count?
Measurement end
Count = 3
Measurement end
Trigger application
Measurement
value acquisition
Count = 3
Delay
Sense Aperture
Measurement end
Count = 3
Sense Aperture
PLZ-5W