Performance Verification
Performance
Check/Adjustment Files
Table 4-6: Waveforms and sequences in performance check disk
No.
File name
1
MODE.WFM
2
PULSE.WFM
3
SINE.WFM
4
DOUT.PAT
5
AMP1.SEQ
6
AMP2.SEQ
7
DC_P.WFM
1
(AMPx.SEQ)
8
DC_M.WFM
1
(AMPx.SEQ)
9
DC0.WFM
(AMP2.SEQ)
10
OFFSET.WFM
4-16
NOTE. The floppy disk file list displayed on the screen does not automatically
update when you replace the diskette with another one. To update the file list,
re-select the floppy disk drive.
Table 4–6 lists the sequence and waveform files on the Performance Check/Ad-
justment disk (063-2983-XX) that are used in these performance tests, the
AWG500–Series Waveform Generator front-panel settings that each file sets up,
and the performance test that uses each file.
NOTE. The files on the Performance Check disk are locked (the files are marked
by the icon
unless the lock is opened.
EDIT menu
SETUP menu
Form
Points
Clock
1000
1 GHz
1000
100 MHz
256
1 GHz
512
100 MHz
1000
100 MHz
1000
100 MHz
1000
100 MHz
1000
100 MHz
in the file list), so the data in these files cannot be changed
Filter
Ampl
Offset
Through
1 V
0 V
Through
1 V
0 V
Through
1 V
0 V
Through
1 V
0 V
Through
1 V
0 V
Through
1 V
0 V
Through
1 V
0 V
Through
20 mV 0 V
Marker setup
Usage
Marker1,2:
Run mode,Trigger
0 to 499: High,
level, Marker
500 to 999: Low
Pulse amplitude, In
ternal trigger accuracy
Sine characteristics
Digital data out
Amplitude accuracy
(Normal out)
Amplitude accuracy
(Direct out)
Amplitude accuracy
Amplitude accuracy
Amplitude accuracy
Offset accuracy
AWG510 & AWG520 Service Manual