9 Diagnosis and service
52
Code
Cause
Text mes-
sage
F025
l Index markers are not
continuously rising, for
Error in the li-
examle unlogical value
nearization
pairs
table
F036
l Failed or interrupted soft-
ware update
No operable
software
F040
l Hardware defect
Error in the
electronics
F080
l General software error
F105
l The instrument is still in the
start phase, the measured
Determine
value could not yet be
measured va-
determined
lue
F113
l EMC interferences
Communicati-
l Transmission error with the
on error
external communication
with 4-wire power supply
unit
F125
l Temperature of the elect-
ronics in the non-specified
Unpermissible
section
electronics
temperature
F260
l Error in the calibration car-
ried out in the factory
Error in the
calibration
l Error in the EEPROM
F261
l Error during setup
Error in the
l False signal suppression
configuration
faulty
l Error when carrying out a
reset
F264
l Adjustment not within the
vessel height/measuring
Installation/S-
range
etup error
l Max. measuring range of
the instrument not suffici-
ent
Removal
l Check linearization table
l Delete table/Create new
l Repeat software update
l Check electronics version
l Exchange of the electro-
nics
l Send instrument for repair
l Exchange of the electro-
nics
l Send instrument for repair
l Separate operating voltage
briefly
l Wait for the warm-up phase
l Duration depending on the
version and parameter ad-
justment up to approxima-
tely 3 min.
l Remove EMC influences
l Exchange 4-wire power
supply unit or electronics
l Check ambient temperatu-
re
l Isolate electronics
l Use instrument with higher
temperature range
l Exchange of the electro-
nics
l Send instrument for repair
l Repeat setup
l Repeat reset
l Check or correct installa-
tion and/or parameter ad-
justment
l Use an instrument with
bigger measuring range
VEGAPULS 62 • Foundation Fieldbus