CAUTION
1. Do not wet the unit or allow foreign matter to get inside
If the unit gets wet, a short circuit may result that could damage the unit or make it extremely hot.
Similarly, a short may occur if foreign matter gets inside. If you accidentally spill a liquid on the unit,
immediately turn off the power supply power switch (set to " ") and unplug the power cord. (Take care
not to touch the cord with wet hands at this time.) After unplugging the unit, use a dry cloth to wipe away
any moisture.
Check if the output connector or AC inlet on the power supply rear panel are wet.
If any liquid or foreign matter gets inside, do not use the unit, but contact your nearest Nikon
representative.
2. Mercury lamp life expectancy
The average life expectancy of an OSRAM Hg lamp HBO
50W/ACL1 is 100 hours. Exceeding 100 hours of use may
lead to an accident such as burst of the lamp. Always
replace with a new lamp once the RUN TIME counter
reading exceeds "10000.0". (Hour meter reads 10000.0
which equals 100 hrs.) After replacing a lamp, be sure to
reset the RUN TIME counter display to "000.0" by pressing
the reset button.
(The RUN TIME counter shows the total accumulated time
that the lamp has been lit. Do not press the reset button
except when replacing with a new lamp. Doing so will
cause you to lose track of the accumulated time of
operation.)
If the lamp bulb becomes noticeably blackened, replace
with a new lamp regardless of the accumulated time of
operation.
Safety Precautions
E-7
Replace with a new lamp when
reading exceeds "10000.0"
Reset button
Power Supply Front Panel